Abstract Background The process of assuring the safety of medical devices is constrained by reliance on voluntary reporting of adverse events. We evaluated a strategy of prospective, active surveillance of a national clinical registry to monitor the safety of an implantable vascular-closure device that had a suspected association with increased adverse events after percutaneous coronary intervention (PCI). Methods We used an integrated clinical-data surveillance system to conduct a prospective, propensity-matched analysis of the safety of the Mynx vascular-closure device, as compared with alternative approved vascular-closure devices, with data from the CathPCI Registry of the National Cardiovascular Data Registry. The primary outcome was any vascular complication, which was a composite of access-site bleeding, access-site hematoma, retroperitoneal bleeding, or any vascular complication requiring intervention. Secondary safety end points were access-site bleeding requiring treatment and postprocedural blood transfusion. Results We analyzed data from 73,124 patients who had received Mynx devices after PCI procedures with femoral access from January 1, 2011, to September 30, 2013. The Mynx device was associated with a significantly greater risk of any vascular complication than were alternative vascular-closure devices (absolute risk, 1.2% vs. 0.8%; relative risk, 1.59; 95% confidence interval [CI], 1.42 to 1.78; P
New England Journal of Medicine
2017
Resnic FS, Majithia A, Marinac-Dabic D, Robbins S, Ssemaganda H et al